Pore Size vs Bubble Point Pressure: ePTFE Reference Data Table

What is the relationship between pore size and bubble point?

Under the same wetting and measurement conditions, an ePTFE membrane with a smaller limiting through-pore generally needs more pressure to displace the wetting liquid. Chaoyue’s catalog lists P200 at 0.10 µm and 200 kPa, P120 at 0.22 µm and 120–150 kPa, P80 at 0.45 µm and 70–100 kPa, and P40 at 1.00 µm and 40–60 kPa.

  • Use the four published pressure entries to shortlist the grade by product code.
  • Compare values only after aligning the wetting liquid, temperature, sample format and bubble point endpoint.
  • Obtain a written release limit and lot results before accepting the catalog number as a purchase specification.
  • Specify filtration performance and operating pressure drop separately for the intended media construction.

What does the ePTFE reference data table show?

The reference data show a higher listed bubble point for the finer pore-size labels in this product family. The relationship is useful for shortlisting these four grades, and the published ranges apply to their respective product codes.

Start from the bubble point membrane range to identify the product family, then use the grade table for the selected ePTFE bubble point filtration media. Expanded polytetrafluoroethylene, or ePTFE, is the membrane material. The source labels one column “pore size” and does not define it as a mean flow value, a largest limiting pore or a particle retention rating. The table below preserves the published designation and converts the pressures to psi for comparison.

Chaoyue catalog values for ePTFE bubble point media
Grade Listed pore size (µm) Listed bubble point (kPa) Approx. bubble point (psi) Listed width (mm) What to verify
P200 0.10 200 29.0 ≤1,400 Confirm whether the single pressure value is typical, a minimum or another specification.
P120 0.22 120–150 17.4–21.8 ≤1,400 Ask what the pore-size label represents and whether both pressure bounds govern lot acceptance.
P80 0.45 70–100 10.2–14.5 ≤1,400 Obtain the method for the pressure band and separate performance data at the intended operating point.
P40 1.00 40–60 5.8–8.7 ≤1,400 Check the lower pressure range against the resolution of the receiving instrument.

Keep P200 at the published single value of 200 kPa until a controlled datasheet identifies its meaning. For P120, P80 and P40, obtain written confirmation of how the listed bands are used for quality release. The psi values use 1 psi ≈ 6.895 kPa and are rounded to one decimal place. Use the agreed reporting unit and unrounded source data when applying a contract limit.

White Chaoyue ePTFE bubble point filtration membrane rolls with a porous structure inset
Chaoyue ePTFE filtration media in roll form. The catalog image shows the form of supply.

The common width entry is an upper catalog width. A converter still needs the required slit width and tolerance, roll length, splice policy and which web areas qualify as usable. Confirm those details together with the grade code because a roll that meets the pressure requirement can still be unusable on a specified slitting or pleating line.

Why does smaller pore size usually mean higher bubble point pressure?

Bubble point is the pressure difference at which gas displaces the wetting liquid from an early through-path and produces the defined test endpoint. Other through-paths can remain filled at that pressure. ASTM F316 describes a visual endpoint based on sustained bubbling and treats the result as evidence about limiting pore size or possible test leakage. ASTM F316 scope and significance.

In the ideal cylindrical capillary model, the pressure difference obeys ΔP = 4γ cos θ ÷ d. Here γ is liquid surface tension, θ is the contact angle at the surface, and d is capillary diameter. For the same wetting behavior and geometry, halving that diameter would double the pressure in the model. A real membrane has paths with varying geometry, and a porometer may apply a shape factor when it converts pressure into an equivalent diameter. Anton Paar capillary flow porometry principles.

Use the capillary equation to understand the trend, and use a documented method to report an equivalent pore diameter. The catalog pore-size labels do not supply the wetting inputs or conversion model needed to solve the equation for these particular grades. For instance, the numerical midpoint of P120’s pressure band is 135 kPa. It is arithmetic, and the table does not give a measured lot average or release target at that value.

Mean flow pore size answers a different question about the contribution of open paths to gas flow. One common porometry method finds where the wet flow curve intersects the half-dry flow curve and converts that pressure to an equivalent size. A full flow-based distribution can help distinguish two samples with similar bubble points but different populations of smaller passages. If a buying specification needs both descriptors, give each its own method, limit and unit.

For supplier comparisons, pair bubble point with the flow data relevant to your application. Two ePTFE samples can open their first through-path at a similar pressure and still pass very different amounts of gas at the same operating pressure. The number of available paths and the resistance of those paths influence flow across the exposed area. For a dry air filtration component, request clean-air resistance at your specified face velocity, which is the volumetric flow divided by the tested face area. If the vendor supplies only a flow number, obtain the corresponding area, pressure and construction before using it to size the filter.

Which test conditions make two results comparable?

Compare bubble point results from the same membrane format, with an agreed wetting liquid, temperature, effective test area and detection endpoint. Matching pressure units alone leaves several causes of a measurement shift unresolved.

The public product table gives pressure values without identifying the liquid, test temperature, fixture or endpoint. Ask for the liquid identity and concentration, wetting and conditioning records, and the differential pressure convention used. Ensure the test medium wets the actual ePTFE surface consistently. With hydrophobic media, a liquid that beads or leaves unwetted paths can produce early gas passage that looks like a larger pore. Different liquid formulations can also produce different pressures even when both wet the membrane. Malvern Panalytical guidance on wetting liquids and porometry.

Specify what the instrument records as the bubble point. The first sustained visual bubbles and a software-detected transition can produce different reported values if the detection criteria, gas flow thresholds or pressure programs differ. Request the instrument model and method revision and, when available, the wet pressure–flow trace with the reported endpoint marked. Record the exposed area and specimen support, especially if a previously approved coupon is moving to a different fixture. Anton Paar guidance on pressurization, sensing and specimen holders.

Treat an unexpectedly low bubble point as a finding to investigate before reclassifying a roll. Incomplete wetting, leakage around the seal or an actual membrane defect can each allow gas through early. Compare the sample with an agreed reference coupon and review the same-lot readings and raw trace. If the discrepancy persists under the agreed method, isolate the lot and follow the supplier quality process. In a disagreement between labs, exchange traceable pieces from the same material and record the fixtures and endpoint criteria used in each run. ASTM F316 discussion of damaged membranes and leaks.

How should manufacturers use the data for incoming inspection?

Set one controlled incoming specification per grade and construction before releasing a purchase order. Identify the membrane code, form, method revision, release rule and the results the supplier must report for each lot. If the supplier uses a different instrument, agree on a comparison using material from the same roll before implementing the receiving test.

For example, 110 kPa falls between the published P80 and P120 bands. A result in that space provides neither a new grade nor the cause of the discrepancy. Check the ordered code, roll identification, method and status of the two catalog bounds before you make a disposition. Apply the signed specification to a production lot even when a public page shows a broader band.

Keep bubble point separate from service pressure drop and particle retention when approving filtration media. Pore opening in a wetted specimen, flow resistance at an operating point and capture under a defined test each require their own data. Specify the intended fluid, flow or face velocity and effective area for performance comparisons. If one membrane meets a pressure target but misses the allowable resistance at the specified flow, compare the permeability or operating curve among eligible constructions. ASTM notes the limits of using capillary-equivalent pore measurements to infer particulate retention. ASTM F316 guidance on particle retention.

Approve incoming ePTFE media and finished filter construction against their respective specifications. A laminate, pleat or edge seal changes what the measured specimen represents. A finished article can include bypass paths at seals or joins that are absent from a flat coupon. Request data for the geometry you buy. When qualifying a roll for conversion, keep a signed reference sample and document which side faces the process stream if the media construction is asymmetric.

A supplier certificate of analysis should identify the lot, the tested grade, the result, the unit and the applicable method revision. Agree how samples are selected if cross-web consistency matters. Identified edge and center locations, or specimens taken from agreed points along a long roll, give the receiving team a way to trace a low reading to where the material was taken. Ask for changes in the membrane construction or test method to be communicated before shipment so the receiving limit can be reviewed against the new conditions.

Agree how to handle a result close to an ePTFE acceptance limit before the first delivery. The reporting precision, measurement uncertainty and repeat-test rule can change the disposition of the same lot. For example, a hypothetical 119.8 kPa reading rounds to 120 kPa at whole-kPa precision. If a signed contract uses a 120 kPa minimum, the agreed precision and decision rule determine how the receiving lab reports and assesses that result. Define when to use a fresh specimen, who reviews a disputed value and which records accompany the decision. Save the instrument reading along with the reported value.

ePTFE bubble point FAQs

What is the bubble point pressure of the 0.22 µm ePTFE grade?

Chaoyue lists 120–150 kPa for P120, its 0.22 µm grade. The applicable test conditions and release limits should come from the current grade specification.

Is P200 specified at a minimum of 200 kPa?

The public table lists a single value of 200 kPa for P200. Ask the supplier whether that number is a minimum, a typical value or another specification before using it as a purchase limit.

Is bubble point the same as mean flow pore size?

Bubble point reflects the opening of an early through-path in a wetted sample. Mean flow pore size is a separate result derived from wet and dry flow curves under a defined porometry method.

Can I compare two results reported in kPa if the liquids differ?

First confirm that the two methods have been correlated for the same membrane. A change in surface tension or contact angle can shift the pressure on the same sample.

Can I apply a raw ePTFE roll specification to a pleated filter?

Use the roll data for incoming membrane selection. Establish a separate acceptance criterion for the finished construction because converted media and assembly interfaces can affect the result.

Why might one laboratory report an unexpectedly low bubble point?

Check the wetting record, fixture seal and endpoint before attributing the shift to the material. Repeat the test under the agreed method with traceable specimens and review the raw data.

Get the grade-specific report before ordering

Send Chaoyue the grade code, raw membrane or converted form, required width and the acceptance method used by your lab. Use the contact page to request full test data, including the pore-size definition, wetting and measurement conditions, lot results and signed release limits. If you are changing suppliers, include the current media code and one representative evaluation record so both labs can identify any differences before an equivalence trial.